This measuring instrument can measure pressure to grasp body pressure, the pressure of clothes, a thing
A measurement method and a proofreading method
Air-pack type contact surface pressure measuring system
Catalogue and detailed specifications.(pdf:A4×4)

1.The basic measurement...The pressure of clothes
2.The basic measurement...Body pressure
3.The basic measurement・・・Pressure to hold
    Measurement operation Manual
    AMI3037-2
     AMI3037-5S,10
    A0101
    Calibration manual
    Handling of Air-pack sensor
Pressure Measuring System for Stocking and Bandages
     AMI3037-SB or A0905
    1.Problem of method of examining pressure of present stockings
 Data of a different stocking from the dummy of a leg
2.The influences of the strength of pressure differ around.<Data example>

Pressure of the skin measures influence on bloodstream
*The pressure that the pressure influence is not in proportion to strength becomes the wound with 1/3.<Data example>
・SKIN PRESSURE/BLOOD CURRENT MEASURING CHAMBER
・CONTACT SURFACE PRESSERE・BLOOD FLOW SENSOR

・Blood vessel organization of skin pressure measuring
A use and data analysis of contact pressure / a bloodstream measuring system
Pressure and an expression of correlation of bloodstream
・Data analysis example:
 
With the same body pressure value, influence on skin is different
data 1: At an angle of around 30 degrees of a bed, I watched body pressure and bloodstream influence

data 2: A bed angle, body pressure, skin bloodstream, the consecutive measurement of skin temperature
data 3: Even if body pressure is the same, influence on skin is different
A system example according to a measurement purpose

A list of presentation of the results of the study
A list of measuring instrument users

JAPANESE ENGLISH

《Principle》

Air packed in an air-bag is put between two contacted matrials,and air pressure generated innside that air-bag is measured.
Air pressure indicated represents the force of contact surface from which the force absorbed by ductility of the material has been deducted.
The correctness of that measured value can be certified by the water
pressure calibration method.
By changing water temperature during calibration, temperature dependence
of the system can also be confirmed. The error has been found
to be extremely small, namely, within 10kPa at45℃.
*The reason why calibration is made by water
The surface of soft materials has a ductility and their surface shape is
always changing when solid materials are in contact, thereby a
resistance is generated due to staggering between two contacting materials.
Such resistance does, on some occasions, not have a reproducibility
in data due to friction, adhering force produced by moisture, or to statical electricity.
Liquid is therefore used for the calibration because it does not produce
a contact pressure and is always subject to an equal force.

<PATENT>


US PAT:NO,5922966 July 13,1999
JAP PAT:NO,3268601 January 18,2002

BY AMI Techno CO.,LTD. /Yukiya KOMINAMI